5. Place the tube weight  on the suction tube
depending on the location where it will be used and
tighten the union nut. The tube weight improves the
stability of the suction tube.
6. Ensure a horizontal and non-slip stand of the
instrument.
NOTE!
Possible loss of the hose weight during use.
`
`
Screw on the hose weight tightly with the cap nut.
`
`
Verify that the hose weight is attached correctly
before use.
Safe sampling
WARNING!
Possible sparking due to electrostatic discharge.
Danger of explosion if combustible liquids are used.
`
`
Connect the earthing terminal  before use.
`
`
Wear the necessary protective clothing.
WARNING!
Bodily injury due to escaping media when over-
filling the sample bottle. Liquids can be aspirated
into the piston and can escape at the back of the
instrument.
`
`
Do not overfill the sample bottle.
`
`
Wear the appropriate protective clothing.
Troubleshooting
Problem
Pump is not generating a
negative pressure
Accessories
Type
Sample bottle (glass, GL45 thread)
Sample bottle (glass, GL45 thread)
Sample bottle (glass, GL45 thread)
Sample bottle (glass, GL45 thread)
Suction tube (PA, electrically conductive, black, straight)
Suction tube (PE, transparent, straight)
Suction hose (PE, transparent, on roll)
Tube weight
To eliminate the error
Repeat assembly steps 1 to 4.
Seal or valve faulty (contact dealer or manufacturer).
1. Activate the handle  and pump liquid into the
sample bottle.
2. To complete sampling, press the vacuum push
button .
3. Screw off the sample bottle and seal it.
WARNING!
Bodily injury due to escaping media when screwing
off the sample bottle and pulling the suction hose
out of the instrument.
`
`
Dismount the instrument carefully.
`
`
Wear the appropriate protective clothing.
After taking the sample
1. Open the clamping screw and remove the suction
tube.
2. Take off the clamping screw and remove the O-ring
and the flat gasket.
3. Rinse the suction tube (together with the hose
weight), O-ring, flat gasket, sample bottle and
instrument with a pH-neutral liquid.
Specification
100 ml
250 ml
500 ml
1000 ml
Ø 6x8 mm, length 220 cm
Ø 6x8 mm, length 200 cm
Ø 6x8 mm, length 1 m
Brass
Number
Item No.
1
5314-0010
1
5314-0025
1
5314-0050
1
5314-0100
10
5305-1024
10
5305-1022
1
8805-0608
1
5305-1028

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

5305-1100

Tabla de contenido