4. OVERVIEW
4.1 Controls
1
Item
Device component
Weighing platform
1
Workflow LED
2
Foot screw
3
Multifunctional display
4
Positioner
5
Connection panel
6
Recessed handle
7
68 •
seca 651/650
4
5
2
3
• Bears the weight of the patient
• For extensions:
seca 464
parts, page 111
Indicates the status of data recording and data transmission (requirement:
Connection to the
seca connect 103
• Illuminated in green: Measurement procedure active
• Flashing green (approx. 5 seconds): Measuring results being sent to the EMR
System (depending on setting)
• Illuminated in green (approx. 5 seconds): Measuring results successfully sent to the
EMR System (depending on setting)
• Illuminated in red (approx. 5 seconds): Error during data transmission or
measurement procedure
NOTE
The data recorded and transmitted are specified in the
software. If you have any queries, contact your administrator or hospital
technician
4 pcs, for precise alignment of the device
Central control and display element
• ➔
Symbols on the multifunctional display (main screen), page 69
• ➔
Symbols on the multifunctional display (menu), page 71
Used to position the multifunctional display on a table or for mounting on a wall
Used for power supply and data transmission ➔
Used for transporting over short distances (2 pcs)
7
6
Description
adapter set required ➔
Optional accessories and spare
software)
Interfaces, page 75
7
seca connect 103