Do not excessively charge the trays, the racks or the baskets. Try always to leave a
free space of 1 or 2 cm between them, in order to let the steam pass through and
also to make drying more easily.

5.7.1 Loose instruments sterilization

Place the instruments over the water resisting paper in the open position and not in
contact with each other.
Please, avoid setting in the same tray different kind of metal instruments.

5.7.2 Containers sterilization

Never set containers hermetically sealed.
Place the containers downwards to avoid water deposits.

5.7.3 Liquids sterilization

Please, set the liquids for sterilization in appropriate containers prepared to support
sterilization temperatures. Place them over the trays, so it could be easier to pick
possible spillage.
The containers should be filled about 2/3 of their capacity. DO NOT CLOSE THEM
hermetically, just cover them with a cotton or some kind of stopper to let the air exit
and not to form any pressure.
Avoid using narrow bottleneck containers.
At the end of the sterilization cycle, let the autoclave freely cool down (till pressure
will be 0 kg/cm²).
DO NOT STERILIZE TUBES NOR WRAPPED MATERIAL
This kind of material is reserved to the autoclaves with a prevacuum purge system.

6 Equipment Description

Micro 8, Med 12 and Med 20 autoclaves are saturated vapour equipments which are
distinguished thanks to the following features:
-
They are not suitable for fixing sterilization units.
-
Suitable for sterilizating:
Unpacked metal load: Clamps, Scalpels, ...
Liquid containers load.
-
Water manual wastepipe.
-
Automatic purge.

6.1 Design features decisive for the equipment life

The maximum service temperature for these equipments is much lower than the one
for which used materials show damage for slow fluids.
-
Fatigue: These equipments have been designed to safety resist the maximum
admissible pressure PS = 2,4 bar, as well as the fatigue in materials due to pres-
sure cycles (from vacuum to maximum pressure).
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: http://www.grupo-selecta.com
February 2020
Tel 34 93 770 08 77 Fax 34 93 770 23 62
Page: 19

Hide quick links:

Publicidad

Tabla de contenido
loading

Productos relacionados para JP Selecta MICRO 8

Este manual también es adecuado para:

Med 12Med 20400174540017564001757

Tabla de contenido