Inspection and maintenance
In this chapter we only check the appearance of the handle. After inspection, if there is no problem, the handpiece should
be immediately reassembled, installing the sealing ring, LED lamp (if any), light pipe (if any), and nipple in sequence to the
handpiece, and then tighten the nipple clockwise.
• Check the handpiece. If there is still visible stain on the handpiece after cleaning/ disinfection, the entire cleaning/
disinfection process must be repeated.
• Check the handpiece. If it is obviously damaged, smashed, detached, corroded or bent, it must be scrapped and not
allowed to continue to be used.
• Check the handpiece. If the accessories are found to be damaged, please replace it before use. And the new accessories
for replacement must be cleaned, disinfected and dried.
• If the service time (number of times) of the handpiece reaches the specified service life (number of times), please replace
it in time.
Packaging
The disinfected and dried handpieces and their accessories are assembled and quickly packaged in a medical sterilization
bag (or special holder, sterile box).
Notes
a) The package used conforms to ISO 11607;
b) It can withstand high temperature of 138 °C and has sufficient steam permeability;
c) The packaging environment and related tools must be cleaned regularly to ensure cleanliness and prevent the introduction
of contaminants.
d) Avoid contact with parts of different metals when packaging.
Sterilization
Use only the following steam sterilization procedures (fractional pre-vacuum procedure*) for sterilization, and other
sterilization procedures are prohibited:
1.The steam sterilizer complies with EN13060 or is certified according to EN 285 to comply with EN ISO 17665;
2.The highest sterilization temperature is 138 ° C;
3. The sterilization time is at least 4 minutes at a temperature of 132 ° C / 134 ° C and a pressure of 2.0 bar ~ 2.3 bars.
4. Allow a maximum sterilization time of 20 minutes at 134 °C.
Verification of the fundamental suitability of the products for effective steam sterilization was provided by a verified testing
laboratory.
Notes
a) Only products that have been effectively cleaned and disinfected are allowed to be sterilized;
b) Before using the sterilizer for sterilization, read the Instruction Manual provided by the equipment manufacturer and follow
the instructions.
c) Do not use hot air sterilization and radiation sterilization as this may result in damage to the product;
d) Please use the recommended sterilization procedures for sterilization. It is not recommended to sterilize with other
sterilization procedures such as ethylene oxide, formaldehyde and low temperature plasma sterilization. The manufacturer
assumes no responsibility for the procedures that have not been recommended. If you use the sterilization procedures that
have not been recommended, please adhere to related effective standards and verify the suitability and effectiveness.
Storage
1. Store in a clean, dry, ventilated, non-corrosive atmosphere with a relative humidity of 10% to 93%, an atmospheric
pressure of 70KPa to 106KPa, and a temperature of -20°C to +55 °C;
2. After sterilization, the product should be packaged in a medical sterilization bag or a clean sealing container, and stored
in a special storage cabinet. The storage time should not exceed 7 days. If it is exceeded, it should be reprocessed before
use.
Notes
a) The storage environment should be clean and must be disinfected regularly; b)Product storage must be batched and marked
and recorded.