Stereomicroscope With Projection Lens - Optika Italy C-WF Serie Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 32
8.4.3 Stereo microscope with projection lens
1. Use a low power magnification and focus the specimen.
2. Reach the highest magnification available using the
zoom knob and then focus the specimen again.
3. Turn on the live-view on the camera, without changing
the focus on the microscope.
4. While observing the image on the screen/monitor, un-
screw the fixing knob ① of the trinocular port (Fig. 17)
and move up or down the movable part of the adapter
② until the focus is achieved on the camera. (Fig. 18)
5. Lock the fixing knob ①. (Fig. 17)
The proper parfocality adjustment is obtained when the
same focus is reached when looking into the eyepieces
and on the screen/monitor.
Page 12
F ig. 17
F
ig. 17
F ig. 18
F
ig. 18

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

C-wfrC-wfr 1

Tabla de contenido