If the response- and recovery time is slow, or if the output signal does not reach the start value
according to the suggested time frame, the sensor is probably contaminated and shall be replaced.
The graph below shows a slow response- and recovery time of a contaminated sensor.

14. Battery indication "low Batt"

The LED "Low Batt" will light up with a constant light when the batteries are close to empty. When the
LED starts to flash, the batteries are to be

15. Troubleshooting

If the actual detector does not generate an alarm when exposed to gas. Check the following:
- GV-offset and sensor signal due to point 6 and 8
- Alarm threshold settings
- DIP-switches, settings of alarm delay
- That the detector is not in service mode
Storage of the instrument
It is of significant importance to the reliability of the instrument,
that the storage conditions are such, that it is not exposed to substances that can
contaminate or damaged the sensor head of the instrument.
This product is intended for use in the industrial area.
Technical specifications are subject to change
743-600-G1 MCT150 Operating Manual 200304
replaced.