Exceptional Situation
Error indicators
The following symbol will appear on the display when measuring abnormal.
Symbol
Weak signal or
pressure change
suddenly
External strong
disturbance
It appears error
during the
process of
inflating
Abnormal
blood pressure
Low battery
Trouble removal
Problem
No power
No inflation
Err and stop working
Cuff leak
Please contact the distributor if you can't solve the problem, do not disassemble the unit by yourself!
Cause
Wrap the cuff properly.
Remeasure with correct way.
When near cell phone or other high radiant
device , the measurement will be failed.
Keep quite and no chatting when measure.
Wrap the cuff properly.
Make sure that the air plug is properly
inserted in the unit.
Remeasure.
Repeat the measurement aſter relax for 30 mins , if get
unusual readings for 3 times,please contact your doctor.
Replace all the worn batteries with new ones.
Check
Check the battery power
Check the polarity position
Whether the plug insert
Whether the plug broken or leak
Whether move the arm when inflate
Check if chatting when measured
Whether the cuff wrap too loose
Whether the cuff broken
Correction
Cause and solutions
Replace new one
Installation for proper placement
of the batteries polarities
Insert into the air socket tightly
Change a new cuff
Keep the body peaceful
Keep quite when measure
Wrap the cuff tightly
Change a new cuff
17