9. What do I do if...
... no weight is displayed with a load on the scale?
– Is the scale switched on?
– Check the mains unit connection.
0. 0 0
...
does not appear before weighing?
– Press the Start key again – the scale may be loaded with a maximum of only 3 kg/
1.2 kg.
----
...
appears before weighing instead of
– Press the Start key again – the scale may be loaded with a maximum of only 3 kg/
1.2 kg.
... the weight display is flashing?
– The HOLD function has not been activated because the baby is moving around too
much on the scale. Press the HOLD key again.
... one segment is illuminated permanently or not at all?
– That point has a fault. Inform your service agent.
St0P
...
appears in the display?
– Maximum capacity has been exceeded - remove load from scale.
– Switch to the weighing range up to 15 kg if necessary.
EEEEE
...
appears in the display?
– Has the scale been started with the transport securing device in place? Remove the
transport securing device (see section entitled "Removing transport securing device"
on Page 14).
– A load far exceeding the permissible weighing range has been placed on the scale (see
also "Exceeding the weighing range" on Page 17).
– A fault has occurred in the electronics. Inform your service agent.
10. Servicing/recalibration
Have the scale recalibrated by authorised persons in accordance with national legal re-
quirements. The year of initial calibration can be found behind the CE symbol on the rating
plate above the number of the appointed office, 0109 (Hessen Weights & Measures Of-
fice). Recalibration is essential if one or more security marks are damaged.
We recommend having your scale serviced by your local service agent before having it
recalibrated. seca Customer Service will be pleased to give you further assistance.
Model 717A
0. 0 0
?
19
GB