CLEANING
The MST Maestro™ handle must be sterilized in a steam
autoclave prior to each use. The handle and tip should not
be assembled during autoclaving.
CAUTION: Improper or inadequate cleaning may result
in particulate matter adhering to the instrument and
exfoliation into the surgical eld. In addition, the function
and life expectancy may become compromised.
The recommended cleaning process for the Maestro™
system is as follows:
1. DURING THE PROCEDURE
Immediately after use, soak I/A tips and handles in
distilled water.
2. POST PROCEDURE CLEANING
This procedure describes the cleaning of the Maestro™
handle with the tip attached to the handle and the
irrigation sleeve removed. See your irrigation sleeve's
user manual for sterilization instructions.
Part I:
a. Fill a 10cc syringe with distilled water.
b. Connect syringe to the handle's irrigation port and
expel the water.
c. Disconnect the syringe and repeat steps a and b two
more times.
d. Repeat this same process for the handle's aspiration
port.
Part II:
a. Remove the tip.
b. Sonicate the devices in fresh DI water for 3 minutes.
Part III:
a. Reattach the tip.
b. Flush both lumens of the handle with tip with 10cc of
air 3 times each.
c. Dry the devices with clean lint-free cloth
NOTE: To avoid damage due to incidental contact, the
Maestro™ components should be kept apart from other
instruments during cleaning and sterilization.