7. Cleaning
Clean the rubber surface and the housing as required using a domestic cleaning agent or
commercially-available disinfectant. Follow the manufacturer's instructions.
8. What do I do if...
... no weight is displayed when a load is placed on the scale?
– Is the scale switched on?
– Check the batteries.
... the display stays dark?
– Is the plug contact on the scale properly connected?
...
0. 0 0
does not appear before the weighing operation?
– Press the Start key again – there must not be any load on the scale
----
...
appears instead of
– Press the Start key again – there must not be any load on the scale.
... one segment is illuminated constantly or not at all?
– The relevant segment has a fault. Inform Servicing.
... the display shows
– Battery voltage is running low. Change the batteries in the next few days.
batt
...
appears in the display?
– Batteries are empty. Put in new batteries.
St0p
...
appears in the display?
– Maximum load has been exceeded.
... the display flashes?
– Take the load off the scale and press the Start key twice. Wait until
then weigh again.
... the display
– The ambient temperature of the scale is too high or too low. Place the scale in an
ambient temperature between +10 °C and +40 °C. Wait about 15 minutes for the
scale to adapt to the ambient temperature and then weigh again.
E
... the display
and a number appear?
– Press the Start key. The scale will then work normally again. If this is not the case,
inform Servicing.
Model 882/pragma 883
0. 0 0
before weighing?
?
appears?
0. 0 0
appears and
GB
31