Guidance  and  manufacture's  declaration  –  electromagnetic  immunity
The  Examination  Light  ri-­magic®  HPLED  is  intended  for  use  in  the  electromagnetic  environment  specified  
below.  The  customer  or  the  user  of  Examination  Light  ri-­magic®  HPLED  should  assure  that  it  is  used  in  such  
an  environment.
Immunity  test IEC  60601  test  level
Conducted  RF
150  kHz  to  80  MHz
IEC  61000-­4-­6
Outside  ISM  bands
Radiated  RF
IEC  61000-­4-­3
26  MHz  to  2.7  GHz
NOTE  1              At  26  MHz  and  800  MHz,  the  higher  frequency  range  applies.
NOTE  2              These  guidelines  may  not  apply  in  all  situations.  Electromagnetic  propagation  is  affected  by      
absorption  and  reflection  from  structures,  objects  and  people.
a
             The  ISM  (industrial,  scientific  and  medical)  bands  between  150  kHz  and  80  MHz  are  6,765  MHz  to  6,795  
MHz;;13,553  MHz  to  13,567  MHz;;  26,957  MHz  to  27,283  MHz;;  and  40,66  MHz  to  40,70  MHz.
b
             The  compliance  levels  in  the  ISM  frequency  bands  between  150  kHz  and  80  MHz  and  in  the  frequency        
range  80  MHz  to  2,7  GHz  are  intended  to  decrease  the  likelihood  that  mobile/portable  communications  
equipment  could  cause  interference  if  it  is  inadvertently  brought  into  patient  areas.  For  this  reason,  an  
additional  factor  of  10/3  has  been  incorporated  into  the  formulae  used  in  calculating  the  recommended  
separation  distance  for  transmitters  in  these  frequency  ranges.
c                        
Field  strengths  from  fixed  transmitters,  such  as  base  stations  for  radio  (cellular/cordless)  telephones  
and  land  mobile  radios,  amateur  radio,  AM  and  FM  radio  broadcast  and  TV  broadcast  cannot  be  
predicted  theoretically  with  accuracy.  To  assess  the  electromagnetic  environment  due  to  fixed  RF  
transmitters,  an  electromagnetic  site  survey  should  be  considered.  If  the  measured  field  strength  in  the  
location  in  which  the  Examination  Light  ri-­magic®  HPLED  is  used  exceeds  the  applicable  RF  
compliance  level  above,  the  Examination  Light  ri-­magic®  HPLED  should  be  observed  to  verify  normal  
operation.  If  abnormal  performance  is  observed,  additional  measures  may  be  necessary,  such  as  
reorienting  or  relocating  the  Examination  Light  ri-­magic®  HPLED.  
d
             Over  the  frequency  range  150  kHz  to  80  MHz,  field  strengths  should  be  less  than  10  V/m.
14
Compliance  
level
3  V
rms
a
3  Vrms
3  V/m
3V/m
Electromagnetic  environment  -­  guidance
Portable  and  mobile  RF  communications  
equipment  should  be  used  no  closer  to  any  part  
 
of  the
Examination  Light  ri-­magic®  HPLED,  
including  cables,  than  the  recommended  
separation  distance  calculated  from  the  equation  
applicable  to  the  frequency  of  the  transmitter.
Recommended  separation  distance
26  MHz  to  800  MHz
800  MHz  to  2.7  GHz
Where  P  is  the  maximum  output  power  rating  of  
the  transmitter  in  watts  (W)  according  to  the  
transmitter  manufacturer  and  d  is  the  
recommended  separation  distance  in  metres  
(m).
Field  strengths  from  fixed  RF  transmitters,  as  
determined  by  an  electromagnetic  site  survey,
should  be  less  than  the  compliance  level  in  each  
b
frequency  range.
Interference  may  occur  in  the  vicinity  of  
equipment  marked  with  the  following  symbol:
a
 