This module can be used to reprocess machine-reprocessable labor-
atory glassware and utensils in a Miele washer-disinfector for laborat-
ory glassware and utensils. Follow the operating instructions for your
washer-disinfector as well as the instructions of the glassware and
utensil manufacturer on how to reprocess their items by machine.
The A 312 injector module is designed for reprocessing glass capil-
lary viscometers with the following dimensions:
- Height
- Capillary tube inner diameter
- Filling tube inner diameter
- Centre of capillary tube to centre of filling
tube
The module has 8 nozzles for reprocessing 8 capillary viscometers.
Up to 2 injector modules can be used in one A 150 lower basket. A
maximum of 16 capillary viscometers can be reprocessed per cycle.
No upper baskets can be used when the A 312 injector module is
used.
en - Appropriate use
Max. 310 mm
Min.  3.0 mm
Min.  5.0 mm
Approx. 20-25 mm