Accu-Chek FlexLink Manual Del Usuario página 14

Ocultar thumbs Ver también para FlexLink:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 56
that no one can injure themselves on the needle
to avoid risk of infection.
Picture 8
Attach the luer-lock connector (K) to the pump
by screwing it firmly into the pump adapter.
Follow the pump manufacturer's User's Manual
to prime the transfer set. Prime the transfer set
until insulin drips from the connector needle tip.
Do not leave air bubbles in either the cartridge or
the transfer set. Connect the transfer set to the
newly inserted head set and fill the empty space
of the cannula housing with a bolus of 1.0 units
(U100 insulin).
WARNING
You must fill the empty space of the newly
inserted head set after connecting the
transfer set. Failure to do so will result in
missed insulin.
If the coupling does not click noticeably and
audibly together, use a new head set and a
new transfer set.
Filling volume of transfer set:
12"/ 30 cm: approx. 6 units* approx. 60 μl
24"/ 60 cm: approx. 11 units* approx. 110 μl
31"/ 80 cm: approx. 14 units* approx. 140 μl
43"/ 110 cm: approx. 18 units* approx. 180 μl
en
11

Publicidad

Tabla de contenido
loading

Productos relacionados para Accu-Chek FlexLink

Tabla de contenido