BASIC PARAMETERS
4. Input each parameter. Press Apply key to save
the input value and setup. The display returns to
the Tools screen.
●Each setting
• Control:
Choose the operation control mode between the NORMAL (normal control) or the ECO (eco control). By
holding the Control slide key and sliding it right, operation control mode is changed to ECO. Factory
setting: NORMAL.
• Compressor Delay:
The time from turning ON this unit until starting its compressor. This unit is required a large electric power
at the moment its compressor starts. When some units are in a same room, set so as to be shifted to
each other the compressor delay times, to prevent the simultaneous start of all compressors after power
failure. Settable range: 3 minutes~15 minutes, factory setting: 3 minutes.
Note: This unit has two compressor A and B. The compressor whose cumulative operating time is shorter
than the other one, starts first, after 1 minute the other starts.
5. Press Top key to return to the Top screen.