Klarstein SUMMERSHINE SLIM Manual De Usuario página 22

Lámpara terapéutica
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

EN
Immunity Test standard
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast transient/burst
IEC 61000-4-4
Surge IEC 61000-4-5
Voltage dips, interruptions
and variations on power
supply input lines
IEC 61000-4-11
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
Note: U
is the a.c. mains voltage prior to application of the test level.
T
22
IEC 60601-1-2 Test Level
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air
± 2 kV for power supply lines
± 1 kV signal input / output
100 kHz repetition frequency
± 0.5 kV, ± 1 kV differential lines
± 0,5 kV, ± 1 kV, ± 1 kV common mode
0 % U
; 0.5 cycle. At 0°, 45°, 90°, 135°,
T
180°, 225°, 270° and 315°
0 % U
; 1 cycle and 70 % U
T
cycles;
Single phase: at 0°.
0 % U
; 250/300 cycle
T
30 A/m
50 Hz/60 Hz
3 V
0.15 MHz – 80 MHz
6 V in ISM and amateur radio bands
between 0.15 MHz and 80 Mhz
80 % AM at 1 kHz
10 V/m
80 MHz – 2.7 GHz
80 % AM at 1 kHz
Compliance Level
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air
± 2 kV for power supply lines
Not applicabel
100 kHz repetition frequency
± 0.5 kV, ± 1 kV differential lines
Not Applicable
0 % U
; 0.5 cycle. At 0°, 45°, 90°,
T
135°, 180°, 225°, 270° and 315°
; 25/30
0 % U
; 1 cycle and 70 % U
T
T
cycles;
Single phase: at 0°.
0 % U
; 250/300 cycle
T
3 A/m
50 Hz/60 Hz
3 V
0.15 MHz – 80 MHz
6 V in ISM and amateur radio bands
between 0.15 MHz and 80 Mhz
80 % AM at 1 kHz
10 V/m
80 MHz – 2.7 GHz
80 % AM at 1 kHz
; 25/30
T