Troubleshooting

COY381-FRL electrolyte and cleaner set
• Order No. 50081036
COY31-PF abrasive foil
• Order No. 51506973
15
Troubleshooting

15.1 Troubleshooting instructions

If any of the following problems occur, test the measuring instrument in the order indicated.
Problem
Sensor in air,
instrument displays 0.0 mg/l or
0 % O2
Sensor cannot be calibrated
Sensor cannot be calibrated
despite having replaced the cap
Incorrect temperature displayed

15.2 Return

If the unit requires repair, please send the measuring instrument and sensor cleaned to your local
sales center.
Please use the original packaging when returning the instrument.
Please enclose the duly completed "Declaration of Contamination" (copy the second last page of
these Operating Instructions) with the packaging and the transportation documents. The in-
strument cannot be repaired if a duly completed Declaration of Contamination is not
provided!

15.3 Disposal

The instrument contains electronic components and must therefore be disposed of in accordance
with regulations on the disposal of electronic waste.
Please observe local regulations.
30
Cause
Sensor not connected correctly
Cable defective
No electrolyte in membrane cap
Buildup on the membrane
Electrolyte used
Buildup on the working electrode
Temperature sensor defective
Remedial action
Connect the sensor correctly
Fill with electrolyte or replace electro-
lyte.
Replace the membrane cap
Replace electrolyte
Clean the working electrode
Send in sensor for repair