Immunity Test
Electrostatic
Discharge (ESD)
IEC 61000-4-2
Electrical fast
Transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
is the a�c� mains voltage prior to application of the test level�
Note: U
T
IEC 60601
Test Level
±8 kV contact
discharges
±2 kV, ±4 kV, ±8
kV, and ±15 kV air
discharges
±2 kV at 100 kHz
repetition rate for
power supply lines
±1 kV for input-
output lines; both at
100 kHz repetition
rate
±1 kV differential
mode
±2 kV common mode
<5% U
(>95% dip in
T
U
) for 0�5 cycle at 45
T
degree increments
<5% U
(>95% dip in
T
U
) for 1 cycle
T
70% U
(30% dip in
T
U
) for 0�5 seconds
T
<5% U
(>95% dip in
T
U
) for 5 seconds
T
30 A/m
SimplyGo Mini User Manual
Compliance
Level
±8 kV contact
discharges
±2 kV, ±4 kV, ±8
kV, and ±15 kV air
discharges
±2 kV at 100 kHz
repetition rate for
power supply lines
N/A - Device does not
have user I/O lines
that are longer than
3m in length�
±1 kV differential
mode
N/A Device is Class 2
and does not have an
earth connection�
<5% U
(>95% dip in
T
U
) for 0�5 cycle at 45
T
degree increments
<5% U
(>95% dip in
T
U
) for 1 cycle
T
70% U
(30% dip in
T
U
) for 0�5 seconds
T
<5% U
(>95% dip in
T
U
) for 5 seconds
T
30 A/m
Electromagnetic
Environment
- Guidance
Floors should be
wood, concrete or
ceramic tile� If floors
are covered with
synthetic material,
the relative humidity
should be at least
35%�
Mains power quality
should be that of
a typical home or
hospital environment�
Mains power quality
should be that of
a typical home or
hospital environment�
Mains power quality
should be that of
a typical home or
hospital environment�
If the user of the
device requires
continued operation
during power mains
interruptions, it
is recommended
that the device be
powered from an
uninterruptible
power supply or a
battery�
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical hospital or
home environment�