MANUAL DE INSTRUCCIONES CODIGO 80241 REV D
J
.P. SELECTA s.a.
Do not excessively charge the trays, the racks or the baskets. Try always to leave a
free space of 1 or 2 cm between them, in order to let the steam pass through and
also to make drying more easily.
5.7.1 Loose instruments sterilization:
Place the instruments over the water resisting paper in the open position and not
in contact with each other.
Please, avoid setting in the same tray different kind of metal instruments.
5.7.2 Containers sterilization:
Never set containers hermetically sealed.
Place the containers downwards to avoid water deposits.
5.7.3 Liquids sterilization:
Please, set the liquids for sterilization in appropriate containers prepared to support
sterilization temperatures. Place them over the trays, so it could be easier to pick
possible spillage.
The containers should be fi lled about 2/3 of their capacity. DO NOT CLOSE THEM
hermetically, just cover them with a cotton or some kind of stopper to let the air
exit and not to form any pressure.
Avoid using narrow bottleneck containers.
At the end of the sterilization cycle, let the autoclave freely cool down (till pressu-
re will be 0 kg/cm²).
DO NOT STERILIZE TUBES NOR WRAPPED MATERIAL
This kind of material is reserved to the autoclaves with a prevacuum purge system.
6 Equipment description
The Micro 8 autoclave is a saturated vapour equipment which is distinguished
thanks to the following features:
-
They are not suitable for fi xing sterilization units.
-
Suitable for sterilizating:
Unpacked metal load: Clamps, Scalpels, ...
Liquid containers load.
-
Water manual wastepipe.
-
Manual purge.
6.1 Design features decisive for the equipment life
The maximum service temperature for this equipment is much lower than the one
for which used materials show damage for slow fl uids.
-
Fatigue: This equipment has been designed to safety resist the maximum admis-
sible pressure PS = 2,4 bar, as well as the fatigue in materials due to pressure
cycles (from vacuum to maximum pressure). The theoretical
Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: http://www.grupo-selecta.com
Junio 2012
(Sujetas a modifi caciones sin previo aviso)
Tel (34) 93 770 08 77 Fax (34) 93 770 23 62
Pag.: 18