2.
Install a new set of "AA" batteries into the foot
control (see section 7.9). Leave the battery cover of
the Tap-On
Foot Pedal open so the red push button
TM
is accessible.
3.
Maintain a distance of no more than 10 feet (3
meters) between the base unit and Tap-On
Pedal during the synchronization process.
4.
Remove any inserts from the handpiece and adjust
the Power Level Control out of Rinse Mode. Turn the
Main Power switch to the ON (I) position and wait for
the Diagnostic Display graphics to light (refer to
Section 8.2).
5.
While all graphics are lit, press the Purge button,
located on the Diagnostic Display.
PURGE
The graphics will begin to blink in a sequential
pattern, representing the synchronization mode.
This mode will last 5 to 6 seconds.
6.
During this mode, press the red button located in the
battery compartment of the Tap-On
will complete the synchronization process.
7.
Synchronization is successful when all graphic lights
blink at the same time.
8.
To verify proper communication, press the foot
control to the Boost position (Tap-On
fully depressed – 2
nd
graphic on base unit illuminates.
9.
Attach battery cover and tighten the screw.
10.
In the event communication cannot be established,
temporarily use the supplied Auxillary Tap-On
Pedal Cable to connect the Tap-On
directly to the unit.
Foot
TM
Foot Pedal. This
TM
Foot Pedal
TM
position) and ensure the Boost
TM
Foot Pedal
TM
Foot
12