5.
Maintenance
!
Work on electrical equipment is only to be
performed by persons with sufficient knowledge and
experience of electrical systems!
5.1
Cleaning
!
Do not use abrasive cleaning agents and coarse
workshop cloths!
¶
The housing and the LCD are only to be cleaned
using a soft cloth and neutral cleaning agents.
5.2
Calibration
For applications demanding mandatory calibration,
regular calibration is essential. (depending on the
regulations of the respective country).
5.3
Service intervals
i
Routine maintenance ensures constant readiness for
operation.
The following maintenance intervals are to be observed
unless otherwise specified by the approval authorities.
5.3.1
Half-yearly maintenance
¶
Replacement of coarse filter GF1 (fig. 4, pos. 8) in
sampling hose.
¶
Replacement of coarse filter GF2 (fig. 4, pos. 1),
refer to Section 5.4.4.
¶
Check whether both PVC hoses are connected to the
gas outlets (fig. 4, pos. 6 and 7).
¶
Visual inspection of sampling probe (fig. 4, pos. 10).
¶
Sampling probe leak test (refer to Section 5.4.1).
5.3.2
Annual maintenance
This work is to be performed by a qualified
maintenance service. It includes the half-yearly
maintenance and the following additional items:
¶
Checking of BEA 055 measurement accuracy with
a test gas.
¶
Replacement of activated charcoal filter
(fig. 4, pos. 4) in zero gas path.
¶
Replacement of coarse filter GF3
(refer to Section 5.4.4)
Compliance with the specifications of the relevant
authorities must be ensured.
Robert Bosch GmbH
11-16V
20 W
459907_4Ko.eps
11
Fig. 3:
1 Coarse filter GF2
2 Coarse filter GF3
3 8 m hose
4 Activated charcoal filter AF1
5 O
6 Measurement gas outlet (transparent PVC hose)
7 Gas outlet and condensate outlet (transparent PVC hose)
8 Coarse filter GF1
9 30 cm Viton hose (black)
10 Exhaust gas sampling probe
11 Plastic hose for leak test
5.4
5.4.1
To ensure accurate exhaust emissions tests, it is
essential that there are no leaks in the sampling system
(exhaust gas sampling probe, hoses, coarse filters). It is
therefore advisable to conduct the leak test daily.
5.4.2
Keep the opening at the tip of the exhaust gas
sampling probe clean. In the event of HC residue and
condensate, detach the exhaust gas sampling probe
from the hose and blow compressed air through it in
the opposite direction to the intake flow.
5.4.3
Check for damage. In the event of HC residue or
condensate, detach the hose from the measuring device
and blow compressed air through it in the opposite
direction to the intake flow.
Maintenance | BEA 055 | 21
BEA 060
GF 3
O2
AF I
N0
7
6
5
4
10
BEA 055
sensor
2
Maintenance
Leak-tightness of sampling system
Exhaust gas sampling probe
Exhaust gas sampling hose
en
GF 2
1
2
3
8
9
|
1 689 989 111
2019-07-24