Sample values table for the "6-channel portable measuring instru-
ment"
The configuration for the "6-channel portable measuring instrument" is
slightly different than the "3/4-channel portable measuring instrument"
With the "6-channel portable measuring instrument", the voltage signal
values of 0 V and 10 V are always entered under SIGNAL for FROM/TO.
This is also the case when the measuring range on the current/voltage
converter is set to 3 V, 10 V or 48 V. The FROM/TO setting for a current
signal under SIGNAL is also always fixed to the values 0 mA and 20 mA.
This is also the case when the current/voltage converter is set to a meas-
uring range of 200 mA or 4 A.
The sample values table and the following example clarify this in more
detail.
Current/voltage converter
Signal
Signal output
setting
±48 V
0...3 V
±10 V
0...3 V
±3 V
0...3 V
±20 mA
0...3 V
±200 mA
0...3 V
±4000 mA
0...3 V
Menu setting on the portable meas-
uring instrument
UNIT
FROM
TO
0
300
0
500
0
700
0
600
10
1000
-1
1
SIGNAL
FROM
TO
0 V
10 V
0 V
10 V
0 V
10 V
0 mA
20 mA
0 mA
20 mA
0 mA
20 mA
4