Service information
Maintenance of a medical device is wholly the responsibility of the owner of that device. Failure to maintain a
device in accordance with the instructions may invalidate the device's warranty. Furthermore, failure to main-
tain a device may compromise the clinical condition or safety of users and/or their caregivers.
EN
1st Day
Read the User Guide carefully before use and
•
store it in a safe place for future reference
•
Store the enclosed tools on/at the product
Before every use
Use a cloth with warm water and a mild deter-
•
gent without chlorine to clean the product
Check that all castors are moving freely
•
and that all wheel locks/brakes are easily
activated
Make sure all cables are securely plugged in
•
before use in the shower
•
Inspect all cables for cracks or signs for worn-
out parts. Never use a product which appears
to be substandard or faulty. Contact your local
dealer for service
•
Disinfect the upholstery. please check the
disinfection section in this User Guide for
more information
•
Inspect the detachable parts. If required,
disinfect the parts with a 70% disinfectant
solution. It is recommended to wipe of any
residue and dirt using a cloth wrung in warm
water and a mild detergent/soap without
chlorine. Let the detachable parts dry before
disinfection
Daily
•
Use a dry cloth to clean the product
Visual inspection. Repair or replace any dam-
•
aged or worn-out parts
•
Ensure that all fasteners are in place and fit-
ted correctly
•
Check all fittings and buckles for signs of worn
out parts
Be aware of any MAX mark indications before
•
adjustment
Weekly
Check that all castors are moving freely and
•
that all wheel locks/brakes are easily activated
•
Wipe off any residue and dirt from the product
using a cloth with warm water and a mild deter-
gent/soap and dry before use
Monthly
All nuts and bolt on this product must be
•
checked and tightened to avoid unnecessary
failures
•
Oil the swing-away parts. We recommend you
use a professional lubricant system
ENGLISH
Annually
Inspect the frame for cracks or signs for worn-
•
out parts and carry out annual service. Never
use a product which appears to be substan-
dard or faulty
Inspect the piston rods are greased, extract the
•
piston rods to check grease
Warewashing machine
This product can be washed at:
85°C using a mild detergent in 3 minutes in a wash-
ing machine designed for medical devices. Use the
machine's drying function to dry the product. Clean
the accessories seperately.
60°C using a mild detergent in 15-20 minutes in a
washing machine designed for medical devices. Use
the machine's drying function to dry the product.
Clean the accessories seperately.
Before washing:
Avoid degreasing of the piston rods, the ac-
1.
tuators should be retracted to minimum stroke
and without load before washing.
Remove all accessories, unplug all cables and
2.
make sure to dismount the battery, control-
box and handset. Leave the actuators with the
cables mounted on the product (see page 24)
After washing:
Ensure correct placement of cables inserted
3.
into the control box when reinstalling the pow-
er unit system (see page 25)
Check that the handset functionalities works
4.
correct
Disinfection
The upholstery can be disinfected with a 70% disin-
fectant solution. It is recommended to wipe off any
residue and dirt from the upholstery, using a cloth
with warm water and a mild detergent/soap without
chlorine and let it dry before disinfection.
Service interval
A detailed inspection of the product must be made
every 12 months (every 6 months for heavy users)
and every time the product is reissued for use. The
inspection should be carried out by a technically
competent person who has been trained in the use
of the product.
The lifetime of this product in normal use is 5 years
if all maintenance and servicing is carried out in ac-
cordance with the manufacturer's instructions and
demonstrably recorded.