NomoLine® ISA™
Mounting Kit
INTENDED USE
The NomoLine® ISA™, Mounting Kit is used to attach the NomoLine ISA Gas Analyzer to a host device or monitor. (Such as the
Masimo® Root® device).
DESCRIPTION
The NomoLine ISA, Mounting Kit is an accessory for use with a host device/monitor and includes:
• (1) Mounting Plate
• (1) Mounting Bracket
• (1) Adhesive Tape
INSTRUCTIONS
Mount the NomoLine ISA gas analyzer to a host device/monitor:
1. Peel off the backing from the adhesive tape.
2. Refer to Fig. 1. Attach the adhesive tape to the mounting plate.
3. Refer to Fig. 2. Slide the bracket into the mounting plate.
4. Peel off the release liner.
5. Refer to Fig. 3. Verify placement of the bracket on the host device/monitor. (Such as the Masimo Root device - shown).
6. Refer to Fig. 4. Attach the assembled bracket and plate to the host device/monitor. (Such as the Masimo Root device -
shown).
7. Refer to Fig. 5. Slide the NomoLine ISA gas analyzer into the mounted bracket.
Mount the NomoLine ISA gas analyzer to a pole clamp:
1. Do not attach the adhesive onto the mounting plate.
2. Refer to Fig. 6. Screw the mounting plate onto the pole clamp using the screws provided.
3. Refer to Fig. 7. Slide the bracket into the mounting plate.
4. Refer to Fig. 8. Attach the clamp onto the pole.
5. Refer to Fig. 5. Slide the NomoLine ISA gas analyzer into the mounted bracket.
WARNINGS
• The adhesive cannot be removed or repositioned once it adheres to the host device/monitor.
• Do not pick up the host device/monitor from the bracket.
• The ISA analyzer must be properly positioned in mounting bracket before use.
• Do not use the ISA analyzer device with mounting accessory if mounting plate or bracket are damaged.
• The mounting plate and bracket must be properly mounted to external equipment before use.
CLEANING
To clean use a soft cloth wipe each of the parts with the following chemicals for 30 seconds.
Cleaning solutions
• 70% isopropyl alcohol
• Glutaraldehyde
• 10% bleach (sodium hypochlorite) to 90% water solution
• Quaternary ammonium chloride
CAUTION
• Do not immerse in any liquid solution.
• Do not sterilize by irradiation, steam, autoclave, or ethylene oxide.
• Do not clean with chemicals not approved above.
ENVIRONMENTAL
Storage Temperature
Relative Humidity
DISPOSAL
Dispose the used product at the end of the expected service life in accordance with local regulations.
WARRANTY
Masimo warrants to the initial buyer only that these products, when used in accordance with the directions provided with
the Products by Masimo, will be free of defects in materials and workmanship for a period of six (6) months. Single use
D I R E C T I O N S F O R U S E
-40°F to 158°F (-40°C to 70°C)
10% to 95% non-condensing
3
en
10451D-eIFU-0521