Specifications; Functional Specifications; Insulation; Continuity - Chauvin Arnoux C.A 6511 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

FUNCTIONAL SPECIFICATIONS

Insulation

Measurement range : 0.1 to 1000 MΩ
Short circuit current
* C.A 6513 only
The time required for the voltage present on the terminals to decrease by 90% of its value,
once the MΩ button is released (= discharge time), is 1s/µF.

Continuity

Measurement range : 0 to 10 Ω with reversal of the measurement current

Resistance (C.A 6513 only)

Measurement range : 0 to 1000 Ω

SPECIFICATIONS

Scale
0,1 MΩ to 1000 MΩ
Accuracy
± 5% of the measurement
Range
Empty voltage
Test current
R ≤ 500 kΩ
Scale
Accuracy
Short circuit
current
Empty voltage
Scale
Accuracy
Short circuit
current
Empty voltage
17
500 V
1000V*
600V
1200V
≥ 1 mA for
≥ 1 mA for
R ≤ 1 MΩ
≤ 6 mA
≤ 6 mA
0 - 10 Ω
± 3% of end of scale
≥ 200 mA
4.5 V ≤ V ≤ 6.5 V
0 - 1000 Ω
± 3% of end of scale
≥ 2mA
4.5 V ≤ V ≤ 6.5 V

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

C.a 6513

Tabla de contenido