9
Technical data
Power consumption
Rated voltage / fuse type
Nominal frequency
Voltage tolerance for all mains voltages
Mains cable
Protective class
Pollution degree
Pressure hose connection
Nominal pressure
Max. operating pressure
Operating temperature of water
Temperature limit
Height
Width
Depth
Weight
10
Information on malfunctions
Error messages
Error
Cause
E 98
Sensor break / short circuit (heating plate) or
automatic software - Safety shutdown due to
high temperature
E 99
Sensor break / short circuit (water bath) or
automatic software safety shutdown due to
high temperature
E 18
Heating defect: If there is no temperature
increase after a defined period of time.
E 19
Heating control: If the control temperature
exceeds the target temperature for a defined
period of time
11
Service
We appreciate your comments, feedback, and suggestions.
Contact in the countries and more information are available by the displayed QR code or at our website
www.kulzer.com
Micro USB port: (See section 5)
The USB port is used for software updates, which may only be performed by trained service personnel.
Therefore, customers are not permitted to connect USB cables/adapters, etc.
There is a risk of damage if not performed by trained service personnel!
Polymerization unit for hot- and cold-curing resins
Remedy
- Press the Home button (11).
- Turn off the device
- Check water level (see section 7.2)
- Allow device to cool for approx. 10 min
- If the error persists, please contact your service partner.
Press the Home button (11).
- Turn off the device
- Check water level (see section 7.2)
- Allow device to cool for approx. 10 min
- If the error persists, please contact your service partner.
- Press the Home button (11).
- Turn off the device
- Check water level (see section 7.2)
- Allow device to cool for approx. 10 min
- If the error persists, please contact your service partner.
- Press the Home button (11).
- Turn off the device
- Check water level (see section 7.2)
- Allow device to cool for approx. 10 min
- If the error persists, please contact your service partner.
Operating Instructions
650 W
100 V, 2 x T8A / 250 V
120 V, 2 x T6.3A / 250 V
230 / 240 V, 2 x T4A / 250 V
50 / 60 Hz
+-10%
EU - H05V2V2-F 3G 0.75 250VAC/16A
US - SJT3X18AWGWB105C C13/2,5m black 125VAC/15A
1
2
3 to 10 bars
2 bars
2.8 bars
40 – 100°C (113 – 212°F)
Deviation + / - 2.5 °C (4.5 °F) or approx. 5%
Dimensions
Required space
Height 300 mm
approx. 450 mm
Width 305 mm
approx. 315 mm
Depth 400 mm
approx. 550 mm
Weight 8.6 kg
GB