CAUTION
● Magnetic and electrical fields are capable of interfering with the proper performance of the
CPR sensor. For this reason make sure that all external devices operated in the vicinity of the
CPR sensor comply with the relevant EMC requirements. Mobile phone, X-ray equipment
or MRI devices are a possible source of interference as they may emit higher levels of
electromagnetic radiation.
● Always carry the CPR sensor properly to avoid damage caused by drop, impact, strong
vibration or other mechanical force.
● Dry the CPR sensor immediately in case of rain or water spray.
● At the end of its service life, the CPR sensor, as well as CPR sensor cable, must be disposed of
in compliance with the guidelines regulating the disposal of such products. If you have any
questions concerning disposal, please contact us.
● Dispose of the package material as per the applicable waste control regulations. Keep it out
of children's reach.
Indicators
CPR Sensor Status
Power off
Sleep (only for CPR sensor
configured with a battery)
Normal use
Inactivity
Self-test error
Battery
The battery indicators indicate battery status as follows:
 Green: the battery power is normal.
 Yellow: the battery is being charged.
 Red: the battery power is low.
 Off: the battery is depleted or the CPR sensor is not configured with a battery.
When the CPR sensor configured with a battery is in the sleeping status, the battery indicator blinks
once every 5 seconds in normal case.
When the battery is low, you should charge the battery by connecting the CPR sensor to a Mindray
defibrillator or monitor supporting the CPR function.
Compression Depth/
Recoil Indicator
Off
Off
indicates the current compression
depth,
 Yellow area: 0 to 5 cm
 Green area: 5 to 6 cm
 Red area: 6 to 8 cm
the first indicator lights in blue
the first indicator lights in blue
3
MR6401 CPR Sensor Operator's Manual
Compression Rate Indicator/
Inactivity Timer
Off
Off
indicates the current compression
rate
displays interruption time in
seconds since the last compression
displays EXX (XX represents code
01 to 99)