Changing The Filter - Beurer IH20 Instrucciones Para El Uso

Ocultar thumbs Ver también para IH20:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 47
In the case of oversensitivity of the bronchial system, medicines containing essential
oils can sometimes cause acute bronchiospasm (a sudden, cramp-like tightening of
the bronchi accompanied by breathlessness). Always ask your doctor or pharmacist
for advice!
6
End of inhalation: If the spray is only output sporadically or if the noise changes
during inhalation, you can end the treatment.
• After treatment, turn off the nebuliser at the switch [3] and disconnect it from the
mains.
• After use, replace the atomiser [9] in its holder [7].
7
Clean the device.
For information about this, see "Cleaning and disinfection" on page 26.
7

Changing the filter

Check the filter at regular intervals (e.g. after every 10th use). If the filter is very contami-
nated or blocked, replace the used filter. If the filter has become wet, it must also be re-
placed with a new one.
CAUTION:
• Do not try to clean and re-use the used filter!
• Only use original filters supplied by the manufacturer. Otherwise, your nebu-
liser may become damaged or effective therapy can no longer be guaranteed.
• Never operate the device without a filter.
To replace the filter, proceed as follows:
CAUTION:
• First switch off the unit and unplug it.
• Allow the unit to cool down.
Pull the filter cap [5] forwards and turn it.
1
Note:
If the filter remains in the unit when the cap is re-
moved, remove the filter from the unit using twee-
zers or similar.
Replace the filter cap [5] containing the new filter.
2
Check that the filter cap is firmly in place.
3
25