Memory Storage And Recall; Device Maintenance; Prevent A Malfunction; Technical Specifications - VISIOMED TensioFlash Manual De Usuario

Ocultar thumbs Ver también para TensioFlash:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 19
4. The result will be automatically stored in the current memory bank.
5. Press START again to turn the device off. If you forget, the device will stop automatically after about 1 minute.
Note:
- The value measured will be too low if you hold the cuff higher than heart level, or will be too high if you hold the cuff lower than heart
level.
- During measurement do not speak or move as it may affect the accuracy of the results.
- If you want to stop the reading during the measurement process, just press the START button to stop immediately.
- Consult your doctor if in doubt about using the device.

14. MEMORY STORAGE AND RECALL

1. Press the MEM button to turn on the device.
2. The screen displays the last measurement and reads the results (interpretation of the results). Press the MEM button to review
data in ascending order. The time and date are stored for all results.
3. Press the MEM button to review data in ascending order.
4. To delete measurements from the memory, press the MEM button until 0 is displayed on the screen.
5. Press the START button to turn off the monitor.

15. DEVICE MAINTENANCE

1. Use a soft dry cloth to clean the device. If necessary, use a cloth lightly dampened with tap water.
2. Do not use alcohol, benzene, thinner or other harsh chemicals to clean the device or cuff.
3. Before using please wash your hands. Do not wash or wet the cuff.
4. Avoid placing the device in contact with direct sunlight, high humidity or dust.
5. Fold the pressure cuff and store the device in its box.
Note: If the blood pressure monitor is left unused for long periods, remove the batteries from the device
(otherwise, the batteries may leak and damage the device)

16. PREVENT A MALFUNCTION

- Do not drop the device, avoid jolts and shocks.
- Never insert anything into the device.
- Do not disassemble the device.
- Do not put the device in water.
- Do not crush the pressure cuff.
- The device may not meet its performance specifications if stored or used outside the temperature and humidity ranges specified
22
Visiomed - KD-595 - Manuel d'utilisation - 072017
in the specifications.
- The device retains its safety and performance characteristics for a minimum of 10,000 measurements or three years of use.

17. TECHNICAL SPECIFICATIONS

Product name
TensioFlash - Automatic blood pressure monitor armband
Model
KD-595
Measurement method
Oscillometric system
Classification
Class IIa medical device
Indication
Digital LCD display
Measuring range
Cuff pressure: 0~300 mm Hg.
Systolic: 60~260 mm Hg.
Diastolic: 40~199 mm Hg.
Pulse: 40~180 Beats/min
Accuracy
Pressure +/- 3 mm Hg;
Pulse +/- 5%
Inflation
Inflation by electric pump
Deflation
Pressure released by valve
Memory
Stores 60 last readings.
Power source
4 x 1.5V "AA" alkaline batteries
Battery life
6 months - Approx. 180 readings
Working Temp. &
5~40°C (41~104°F). ≤90% RH
humidity
Storage Temp. &
-20~55°C (-4~131°F). ≤90% RH
humidity
External dimensions
153mm / 108mm / 60 mm (6 1/32" x 4 1/4" x 2 3/8")
Weight
Approx. 290g (Device)
Accessories
Cuff (arm circumference 22~32 cm). Instruction manual, storage case, 4 batteries 1.5V AA
Voltage
6 Volts DC
The manufacturer reserves the right to modify the product technical specifications without prior notice.

18. TROUBLESHOOTING GUIDE

IIf any of the following problems occur when using your TensioFlash®, please refer to this troubleshooting guide. If the problem
persists, please contact our customer service at 0 892 350 334.
Visiomed - KD-595 - Manuel d'utilisation - 072017
23

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Kd-595

Tabla de contenido