Operating; Power Connection - Selecta RE-650 Manual Instrucciones

Reactor para planta piloto
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

4.2   Operating   
Note:
 
Before  starting  the  stirring  function  for  the  first  time,  pay  attention  to  the  relative  position  of  the 
stirring paddle and the bottom of the vessel body to avoid collision.
 
4.2.1
Power Connection
   
When the reactor is connected to the power supply properly, the control panel will enter the welcome 
interface.  5  seconds  later,  it  will  change  to  the  monitoring  interface,  which  displays  the  current 
temperature, stirring speed and torque, as shown in the image. 
 
4.2.2    Stirring speed and time setting 
Press  the  Set  knob  on  the  control  panel  to  enter  the  speed  setting  interface,  turn  the  knob  to  the 
target value (clockwise to increase, counter clockwise to decrease) and then press again the Set knob 
to save the settings. Press it again to enter the time setting interface. Turn the knob slowly to set hour, 
minutes  and  seconds  in  turn,  according  to  the  actual  needs.  After  each  parameter  is  adjusted  to  the  target  value,  they 
need to be saved by pressing the knob. If you don't press the Set knob, it will quit the current setting and return to the 
monitoring interface. 
Note: When the reactor is empty, the stirring speed should be no more than 200rpm. 
 
 
4.2.4 Operation
11.
Feed liquid or solid materials into the reactor vessel slowly through the right port. 
12.
Click button Start/Stop on the control panel to start stirring speed control. 
13.
If the reaction should be carried on under vacuum conditions, start the vacuum pump to pump out the air in the 
vessel until it reaches vacuum state. 
14.
Start the Dynamic Temperature Control System according to actual needs of the reaction, and the temperature in 
the vessel can be controlled automatically when you set it ready. 
15.
When  the  reaction  process  is  finished,  shut  down  the  Dynamic  Temperature  Control  System  when  the 
temperature inside of the reactor is close to room temperature. 
Note: 
After shutting down the temperature control unit, please do not close the valves in the middle of the 
circulation  hoses,  keep  all  the  hoses  clear.  Otherwise,  when  the  temperature  in  the  jacket  or  in  the 
 
vessel changes dramatically, it may cause any crack or broken to the reactor body. 
16.
Shut down the vacuum pump. 
17.
Adjust stirring speed to minimum value; click button Start/Stop to stop stirring. 
18.
Click button Power On/Off to power off the controller. 
19.
Discharge the material from the draining port (14). After emptying the material, please disassemble the draining 
valve and clean the glass part and the PTFE stopper. 
20.
Reinstall the draining valve. 
Note:  When  discharging  the  materials,  slowly  turn  the  black  stopper  at  the  bottom  of  the  discharge  valve 
counterclockwise.  After  cleaning  the  glass  parts,  wipe  the  liquid  on  the  surface  with  a  soft  clean  cloth  to  avoid 
contamination. 
 
 
58
   
 

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

1001840

Tabla de contenido