ProMed PBW- 5.2 Instrucciones De Empleo página 32

Medidor de la tensión arterial para el brazo
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 68
INBETRIEBNAHME
Error code 1 (E1)
appears on the display
Error code 2 (E2)
appears on the display
Error code 3 (E3)
appears on the display
DE
Error code 4 (E4)
appears on the display
EN
FR
IT
Error code 5 (E5)
ES
appears on the display
NL
Error code 6 (E6)
appears on the display
The device keeps
RU
starting the inflating
process.
PL
If the fault cannot be rectified, contact the manufacturer before you try to repair the
device yourself.
SE
FI
The armlet is not put on correctly.
The armlet is not positioned
correctly.
You moved or you moved your
arm during measurement.
The armlet is not connected
properly to the device.
The device does not measure.
An error occurred during
measurement.
The armlet is overinflated.
The batteries are too low.
Switching is blocked.
32
Put on the armlet correctly.
Make sure you do not
move and keep the armlet
at the height of your heart.
.
Make sure you do not
move.
Check if the armlet is
connected properly to the
device.
Due to a very weak or
irregular heartbeat, the
device has difficulties to
determine your blood
pressure.
If this error code
appears more often, we
recommend consulting
your physician to be on
the safe side.
Make sure you do not
move and put the armlet
on again.
If the measured value is
higher than 300 mmHg,
we recommend consulting
your physician to be on
the safe side.
Replace the batteries with
new ones.
Take the batteries out and
insert them again to reset
the device.