Descargar Imprimir esta página

Bürkle UniSampler Ex Instrucciones Para El Uso página 4

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 6
English
General safety requirements
`
`
Read these directions for use carefully before using.
`
`
Keep these directions for use where they will be
accessible for all users.
`
`
Failure to comply with the directions for use shall
render all warranty conditions null and void and
may lead to personal injury and material damage.
`
`
The instrument must only be used by qualified
skilled employees.
Safety instructions before using
`
`
Follow the manufacturer's instructions and safety
instructions of the medium you are sampling.
`
`
Damage to the respiratory tract and other organs
due to off-gassing media. Use in ventilated rooms or
under a suitable exhaust hood.
`
`
When working with explosive substances, follow the
technical rules for combustible liquids (in accordan-
ce with TRBF – (German) Technical Rules on Com-
bustible Liquids) and wear appropriate protective
equipment (explosion zone 0, see annex).
WARNING!
Possible sparking due to electrostatic discharge.
Danger of explosion if combustible liquids are used.
`
`
Connect the earthing terminal  before use.
`
`
Wear the necessary protective clothing.
Safety instructions during and after use
CAUTION!
Liquid entering the instrument due to overfilling
or tipping. Bodily injury due to escaping liquid and
damage to the instrument.
`
`
Shut down and secure the instrument immediately.
`
`
Contact the dealer or manufacturer.
`
`
The instrument must only be used within the
confines of applicable material resistance and for its
designated purpose. Any other type of use will affect
the warranty/guarantee.
`
`
The instrument is not suitable for cleaning in a
dishwasher.
Layout and function
The instrument is made of conductive material and
is suitable for filling non-aggressive, combustible
liquids (group IIA and IIB, see annex).The instrument is
maintenance-free.
Scope of delivery
1 x
Vacuum pump
1 x
Sample bottle with closure, 1000 ml
(glass, GL45 thread)
1 x
Suction hose (PVC, 0,2 m)
1 x
Suction hose (NBR, electrically conducti-
ve, 4.0 m)
1 x
Earthing clamp
1 x
Universal adapter for sample bottles
1 x
Tube weight (brass)
1 x
Bottle stand (PVC)
1 x
Direction for use
1 x
Transport case
`
`
Check delivery for completeness and transport
damage.
`
`
Point out any damage in the packaging to the
freight forwarder/supplier immediately.
Assembly
WARNING!
Possible sparking due to electrostatic discharge.
Danger of explosion if combustible liquids are used.
`
`
Connect the earthing terminal  before use.
`
`
Wear the necessary protective clothing.
1. Insert the sample bottle  into the bottle stand .
2. Attach the adapter for sample bottles  onto the
sample bottle (the rear end of the earthing wire
must be in the bottle).
3. Connect the earthing clamp .
4. Place the free end of the suction hose the liquid to be
sampled.
5. Depending on the sampling location, put the hose
weight  onto the suction hose. The hose weight
improves the stability of the suction hose.
Safe sampling
1. Activate the handle of the vacuum pump and pump
liquid into the sample bottle.
2. To complete sampling, press the valve lever  on
the vacuum pump.
3. Remove the adapter for sample bottles from the
sample bottle and seal it.

Publicidad