CAUTION!
After activating the handle, the pump will continue
to suck liquids. Liquid entering the instrument.
Bodily injury due to escaping liquid and damage to
the instrument.
`
`
Deaerate the pump with the valve lever in good
time.
Troubleshooting
Problem
Pump is not generating a
negative pressure
Accessories
Type
Sampling bottle (Glas, GL 45 thread)
More accessories can be found at www.buerkle.de
Annex
Operational directive in accordance with 1999/92/EC:
Zone 0
Zone 1
Zone 2
Product directive in accordance with 94/9/EC:
Group I
Group II
Group IIA
Group IIB
Group IIC
To eliminate the error
Adapter is not attached properly onto the sample bottle.
Vacuum pump faulty (contact dealer or manufacturer).
a potentially explosive atmosphere is assumed to be continuously present
a potentially explosive atmosphere is assumed to be occasionally present
a potentially explosive atmosphere is not anticipated (and if it does occur, then only
rarely and briefly)
for use in mining
for use in other areas subject to explosion hazard
Maximum permitted gap > 0.9 mm
0.5 mm < maximum permitted gap < 0.9 mm
Maximum permitted gap < 0.5 mm
CAUTION!
Liquid entering the instrument due to overfilling
or tipping. Bodily injury due to escaping liquid and
damage to the instrument.
`
`
Shut down and secure the instrument immedi-
ately.
`
`
Contact the dealer or manufacturer.
After taking the sample
Rinse the suction hose (together with the hose
weight), sample bottle and adapter for sample bottles
with a pH-neutral liquid.