PR Series Balance
InCal Approval Model
Capacity (g)
Readability d (g)
Verification Interval e (g)
Class
Repeatability (STDEV) (g)
Linearity (g)
Stabilization Time Typical (s)
Sensitivity temperature drift
(PPM/K)
Typical Minimum Weight USP
(USP K=2,U=0.10%)
Optimized Min-Weight (g)
(USP, u=0.10%, k=2)
SRP ≤ 0.41d*
Units
Applications
Platform Size (diameter, mm)
Span Calibration Points (g)
Linearity Calibration Points (g)
Tare Range
Power Supply
Assembled Dimensions
(W x D x H) (mm)
Communication
Operating Temperature Range
Storage Temperature Range
Storage Conditions
Net Weight
Shipping Weight
Shipping Dimensions
(W x D x H) (mm)
*SRP refers to the standard deviation for n replicate weightings (n 10).
Table 9-7
SPECIFICATIONS (continued)
PR1602M
PR2202M
1600
2200
0.01
0.01
0.1
0.1
II
II
0.01
0.01
±0.02
±0.02
1
1
±3
±3
20 g
20 g
8.2 g
8.2 g
Basic Weighing, Parts Counting, Percent Weighing, Dynamic Weighing, Density
180
180
1000, 1500,
1000, 2000
1600
0, 1000,
0, 800, 1600
2000
Power input: 100-240V ~ 200mA 50-60Hz 12-18VA
Operating conditions for ordinary lab application: +10°C to 30°C (operability guaranteed
Humidity: maximum relative humidity 80% for temperatures up to 30°C, decreasing linearly to
-10°C to 60°C, humidity 10% to 90%, without condensation
PR4202M
PR5202M
4200
5200
0.01
0.01
0.1
0.1
II
II
0.01
0.01
±0.02
±0.02
1
1
±3
±3
20 g
20 g
8.2 g
8.2 g
g, kg , ct
Determination
180
180
2000, 4000
3000, 5000
0, 2000,
0, 2500,
4000
5000
To capacity by subtraction
Power output 12 VDC 0.5A
209 x 321 x 98
RS232
between +5°C and 40°C).
50% relative humidity at 40°C
7.7 lb / 3.5 kg
11 lb / 5 kg
550 x 385 x 291
EN-39
PR4201M
PR6201M
4200
6200
0.1
0.1
0.1
1
II
II
0.1
0.1
±0.2
±0.2
1
1
±10
±10
200 g
200 g
82 g
82 g
180
180
2000, 4000
5000, 6000
0, 2000,
0, 3000,
4000
6000