4 Calibration and Check

Calibration on white
A calibration on white is necessary every 30 days. If white calibration is ex-
pired, a red cross is displayed. Click on the icon.
1.
Open snap closure and place instrument on white calibration tile.
2.
Click on instrument icon or press button Operate.
3.
First a color measurement on white tile is performed.
4.
It follows a color check on green tile, see below.
5.
Confirm with checkmark in upper right corner.
After successful calibration and color check the red cross turns into a green
checkmark.
Color check on green
A color check on green is recommended every 7 days. This interval is not ob-
served by the instrument and not indicated on the display. Click on the icon.
1.
Open snap closure and place instrument on green color check tile.
2.
Click on instrument icon or press button Operate.
3.
Color measurement on green tile is performed.
4.
Confirm with checkmark in upper right corner.