7309-120-700 Rev-B
Fully charge the power module before first and
every use.
NOTES:
▪
For charging instructions, see the instructions
for use supplied with the battery charger.
▪
For maximum capacity, store the power
module on the battery charger to maintain a
full charge.
▪
Stryker recommends using power modules
that have been charged within the last 5 days.
▪
To install and remove the power module from
a handpiece, see the instructions for use
supplied with the handpiece.
▪
Typical power module run time is 5 minutes or
longer, depending on surgical procedure and
technique.
www.stryker.com
Storage and Handling
CAUTION: ALWAYS transport and store the
equipment within the specified environmental
condition values:
Temperature
Humidity
Limitation
Limitation
Disposal/Recycle
WARNINGS:
▪
Power modules contain Lithium-ion material.
Risk of fire, burns, and explosion. DO NOT
short circuit, open, crush or incinerate.
Consult instructions for use.
EN
Atmospheric
Pressure
Limitation
7