Warranty; Maintenance; Replacing The Battery And The Fuses; Storage - Chauvin Arnoux CA 5003 Manual De Instrucciones

Multímetro
Ocultar thumbs Ver también para CA 5003:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

Our warranty is applicable for three years for the multimeter and one year for
accessories after the date on which the equipment is made available (extract from
our General Conditions of Sale, available on request).
For maintenance, use only specified spare parts. The manufacturer will
not be held responsible for any accident occurring following a repair done
other than by its After Sales Service or approved repairers.

8.1. REPLACING THE BATTERY AND THE FUSES

For your safety, the leads must be disconnected from the multimeter before the
battery cover is opened.
„ To open the cover, turn the screw 1/4 turn, anti-clockwise, using a coin or a
screwdriver.
„ Replace the dead battery by one 9 V battery (type 6F22 or 6LF22).
„ Replace the faulty fuses respecting their value and their type:
- Fuse 1.6 A HPC (6,3 x 32 mm - 500 V - 10 kA)
- Fuse 16 A HPC (10 x 38 mm - 600 V - 10 kA)
„ Refit the cover before using the multimeter.

8.2. STORAGE

If the multimeter is not put into service for a time exceeding 60 days, remove the
batteries and store them separately.

8.3. CLEANING

„ The multimeter must be disconnected from any electrical source.
„ To clean the case, use a cloth slightly moistened with soapy water. Rinse with a
damp cloth. Then, dry rapidly with a cloth or in a hot air stream.