Table 2:Guidance and manufacturer's declaration – electromagnetic immunity
The bosolux auris is intended for use in the electromagnetic environment specifi ed below. The
customer or the user of the bosolux auris should assure that it is used in such an environment.
Immunity test
Electrostatic
Discharge(ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips,
short interrup-
tions and voltage
variations on
power supply
input lines
IEC 61000-4-11
Power frequency
(50/60Hz) mag-
netic fi eld
IEC 61000-4-8
Note: U
is the a.c. mains voltage prior to application of the test level.
T
IEC 60601
Test level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV for input/
output lines
±1 kV lines to
lines
±2 kV lines to
earth
<5% UT
(>95% dip in UT)
For 0,5 cycle
40% UT
(60% dip in UT)
For 5 cycle
70% UT
(30% dip in UT)
For 25 cycle
<5% UT
(>95% dip in UT)
For 5 cycle
3A/m
Compliance level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV for input/
output lines
±1 kV lines to lines
±2 kV lines to earth
<5% UT
(>95% dip in UT)
For 0,5 cycle
40% UT
(60% dip in UT)
For 5 cycle
70% UT
(30% dip in UT)
For 25 cycle
<5% UT
(>95% dip in UT)
For 5 cycle
3A/m
- 40 -
Electromagnetic environment -
guidance
Floors should be wood, concre-
te or ceramic tile. If fl oors are
covered with synthetic material,
the relative humidity should be
at least 30%
Mains power quality should be
that of a typical commercial or
hospital environment
Mains power quality should be
that of a typical commercial or
hospital environment
Mains power quality should be
that of a typical commercial
or hospital environment. If the
user of bosolux auris requires
continued operation during
power mains interruptions, it is
recommended that the bosolux
auris be powered from an
uninterruptible power supply or
a battery.
Power frequency magnetic
fi elds should be at levels cha-
racteristic of a typical location in
a typical commercial or hospital
environment