1.4

Qualification of user

For the intended use, the product may only be operated by med-
ical professionals who have been trained in its use on the basis
of the notes on use.
1.5

Performance characteristics

Performance characteristics relating to the intended use are:
• Transporting HF current from the electrosurgical unit to the
instrument.
2

Maximum Electrical Capacity

The maximum electrical capacity of these products is:
• Number Group 20192-XXX: 5000 V
• Number Group 20196-XXX: 1000 V
3

Safety Instructions

WARNING!
Check insulation for damage before every application.
Erbe recommends carrying out an electrical continuity test be-
fore each application of these products.
If the cable conductor is broken or the electrical transmission of
the cable is interrupted in any other way, this may lead to arc
formation in the patient return line or in the active circuit; this
may cause burns to the patient/surgical staff or start a fire.
If damaged, do not use this product!
Do not use in the presence of combustible or explosive materi-
als!
p