Replacing The Batteries; Getting Started; Buffer Set Selection - ThermoFisher Scientific Elite pH Instrucciones De Operación

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 21
This instruction manual is applicable for Elite pH tester and Elite pH Spear tester.

Replacing the Batteries

The Elite pH tester uses four AAA 1.5 V batteries.
1. To remove the battery cover, See Figure 1. Clear the front catch and
then the back catch, before sliding the cover off.
2. To remove the battery plate, push the center tab towards the
front of the tester as show in Figure 2. Once unlocked, remove the
plate to access the batteries.
3. Invert the tester upside down to remove the batteries. Each side
uses two AAA batteries. Orient each battery with positive terminal
facing downward.
4. To lock the battery plate, align the small taps into the guide ribs
on the housing and then press down. See Figure 4.
Figure 2: Push to unlock

Getting Started

The Elite pH tester has been factory calibrated and usually works well out of the box. However, after extended
periods of non-use, it is best to ensure that the sensor is hydrated. Fill the cap with electrode storage solution or pH
buffer and soak the sensor as needed to get a stable pH reading. A brief rinse with deionized (DI) water is OK, but
avoid soaking or storing in deionized water as this will shorten pH electrode life. Prior to taking the measurements,
periodic calibration with certified standards is recommended for best accuracy.

Buffer Set Selection

Select either USA (pH 4.01, pH 7.00 and pH 10.01) or NIST (pH 4.01, pH 6.86, and pH 9.18) to match your calibra-
tion standard values.
1. Press
to power on.
2. Press
to enter setup window. Press
3. Press
to select Buffer. The display shows USA and NIST.
4. Press
to select USA or scroll down by pressing
5. The display shows the selected buffer with a
Key Info
Scroll
Accept
EXIT
2
Battery Plate Unlock position
Battery Plate Unlock position
To Insert battery plate,
align the small taps into
To Insert battery plate,
the guide ribs on the
align the small taps into
housing
the guide ribs on the
housing
Figure 3: Align tabs
To Insert battery plate,
align the small taps into
the guide ribs on the
housing
to select Settings. The display shows Buffer and Backlight.
Settings
Buffer
Temp Set
Backlight
Reset
Battery Plate locking position
Battery Plate locking position
To lock the battery plate,
press down and slide in
To lock the battery plate,
To Release the battery plate,
press down and slide in
Push away this tap.
To Insert battery plate,
align the small taps into
the guide ribs on the
housing
Figure 4: Push down to lock
Battery Plate Unlock position
To lock the battery plate,
press down and slide in
to toggle between the two buffer standards.
.
USA
USA
NIST
Elite pH and pH Spear Tester Operating Instructions
Figure 1: Removing battery cover
To Release the battery plate,
Push away this tap.
Figure 5: Battery plate unlock position
Battery Plate Unlock position
To lock the battery plate,
press down and slide in
Figure 6: Battery plate lock position
Battery Plate locking position
To Release the battery plate,
Push away this tap.
Battery Plate lock

Publicidad

Tabla de contenido
loading

Productos relacionados para ThermoFisher Scientific Elite pH

Este manual también es adecuado para:

Elite ph spear

Tabla de contenido