Factory Reset
You can reset the pH calibration to the factory reset default by using the factory reset function.
1. Press
to enter setup window. Scroll down by pressing the
display shows User Reset and Fact Reset.
2. Scroll down by pressing the
The display automatically shows No and Yes.
3. Scroll down by pressing
4. Press
to confirm either No or Yes. The display shows the factory reset option with a
Key Info
Scroll
Accept
EXIT
MTS - Manual Temperature Setting (applicable to pH Spear tester only)
The MTS range is 0°C to 60.0°C (32.0°F to140.0°F). User reset will set temperature to default value 25°C or 77°F.
1. Press
to select setup window. Scroll down by pressing
The display shows Set °C/°F and Temp Cal.
2. Scroll down by pressing
3. The lower display shows the current measured temperature reading based on the last set offset and the upper
display shows the current measured temperature reading based on factory default calibration.
4. Dip the tester into a solution of known temperature and allow time for the in built
temperature sensor to stabilize.
5. Press
to adjust the temperature value or press the
value of the solution.
Note: To exit this program without confirming the calibration, press
place.
Key Info
Scroll
Accept
EXIT
HOLD
This feature help to freeze the display reading for a delayed observation.
1. Press
button to freeze the measurement.
PH
pH
4.01
23.5 °C
Elite pH and pH Spear Tester Operating Instructions
to toggle between the resets. Press
to toggle between No and Yes.
Settings
User Reset
Temp Set
Fact Reset
Reset
to toggle between Set °C/°F and Temp Cal. Press
Settings
Set °C/°F
Temp Set
Temp Cal
Reset
PH
pH
4.01
23.5 °C
Press
to
release
to select Reset. Press
to select Fact Reset.
No
Yes
to select Temp Set. Press
to confrim the calibrated value as new temperature
before the automatic confirmation takes
Temp Cal
23.4 °C
23.4 °C
23.4 °C
2. Press
again to release the measurement.
PH
pH
4.01
23.5 °C
Press
to
release
to select Reset. The
.
to select Temp set.
to select Temp Cal.
PH
pH
4.01
23.5 °C
5