Filter Replacement - Medel SMART AEROSOL Manual Del Usuario

Micro-compressor portable nebulizer
Tabla de contenido

Publicidad

EN
5 minutes. Reassemble all parts of the nebulizer and connect
it to the air outlet (Pic. G) and switch on the device for 10-15
minutes.
!
Do not boil or autoclave the air hose the masks.
2. Disinfection:
Cold disinfecting solutions can be used, instead of boiling water,
according to manufacturer's instructions.
NEBULIZER REPLACEMENT
The nebulizer must be replaced after a long period of inactivity,
in case it shows deformations or breakings, or when the nebu-
lizer nozzle is obstructed by dry medicine, dust, etc.. MEDEL
recommends to replace the nebulizer after a period between 6
months and 1 year, depending on the usage.
Use original nebulizers only.
The device is only compatible with the MedelJet Smart Accessories
!
kit.

FILTER REPLACEMENT

In normal usage conditions, the air filter must be re-
placed approximately after 500 working hours or once
a year. Medel recommends to periodically check the air
filter (10 – 12 treatments) and, if the filter shows a grey or brown
colour or is wet, replace it. Remove the filter (Pic. A-4) and re-
place it with a new one. Do not attempt to clean the filter to
reuse it.
!
Only use original filters.
Do not use the device without the filter.
28
100110_MedelSmart_USERmanual_REV.02.21.04.2017.indd 28
07/09/2017 12:32:55

Publicidad

Tabla de contenido
loading

Tabla de contenido