"SILENT POWER"
REPLACING LAMP
a) Silent Power Gold, Silent Power Gold Miniature, Silent Power Silver 4L (Pict. 7)
Disconnect handpiece from faston coupling. Unscrew the front part (9) of the faston coupling and remove bulb.
Put a new bulb in place and screw on part n. 9 again.
b) Silent Power Silver 2, Silent Power Silver 2L (Pict. 8)
Disconnect handpiece from tubing, slide off seal (10) and remove bulb (11). Put a new bulb in, put seal (10) in
position (make sure holes are properly positioned) and connect handpiece to tubing again.
LUBRICATING HIGH SPEED HANDPIECE
Use the spray-can "DAILY OIL" to lubricate at least once daily and in the hereafter indicated
circumstances the high-speed handpiece. Follow the instructions on can, remembering to apply lubricant in
two short sprays. Eliminate any excess lubricant before use by operating the handpiece (hold it over cuspidor)
and wiping in with gauze and cotton. Do this with burr in place to prevent mechanical damage to handpiece.
In addition it is imperative to lubricate as follows:
-Before and after autoclaving
-Before and after every AUTOSTERIL disinfecting or sterilizing cycle.
Regular use of this spray lubricant assures best working order.
Proceed as shown in Picts. 9 and 10, respectively for faston coupling and for screw coupling.
CLEANING, DISINFECTING, STERILIZING
Wear protective gloves! Remove the burr!
Use gauze or cotton soaked in surgical alcohol to clean and/or disinfect handpiece exterior. DO NOT soak
handpiece directly in solution. DO NOT use an ultrasonic cleaner.
Clean the spray nozzle after each operation: use the steel wire supplied and the syringe to blow on the nozzle.
Clean fibre-optic terminals with alcohol-soaked gauze.
In order to maintain a proper level of hygienic safety handpieces must be sterilized in autoclave
with water steam or chemical steam up to 135°C, 210 kPa (2.1 bar) for 20 min, after the use on
each patient.. Never sterilize handpieces in dry-heat sterilizers.
Before and after each autoclaving, lubricate handpiece and clean fibre-optic terminals. Never leave handpieces in
autoclave overnight: always remove them from autoclave after each cycle.
The instrument is supplied not sterile. Before use, sterilize according to the above specifications.
Warning!
Check the autoclave periodically according to the manufacturer's prescriptions!
Temperature exceeding the above stated limit may damage the handpiece!
SERVICING THE CASTELLINI FASTON CONNECTION
Lubricate at least weekly the O-Rings with silicone lubricant S1, distributed by Castellini S.p.A. Wearing protective
disposable gloves, apply a small amount of S1 on fingers and use it for lubricating.
If faston coupling leaks, replace the O-Rings: grasp them with two fingers and slide them along one side and
forward to remove them. Slip on the new rings, fit them properly into the grooves and lubricate.
Important! The faston coupling CANNOT be sterilized in autoclave.
STOP WORKING IMMEDIATELY IN THE EVENT OF
UNUSUAL NOISE, STRONG VIBRATION OR IF BURR IS
IMPORTANT
WE PRESCRIBE TO
NOT KEPT FIRMLY IN POSITION.
Pag. / Seite 11
UK