troubleshooting
... Reading 0 independently of the measured sample.
Broken cable.
Replace the cell.
... Unstable reading.
... Impossible calibration.
Dirty measuring electrodes or air bubbles on their sur-
face.
Clean or shake slightly the cell.
... The obtained readings differ a lot from the real values.
Verify that the conductivity is covered by the measuring
interval of the cell.
Re-calibrate using standards in good conditions.
.... slow cell response
Regenerate the cell.
Specifications
Approx. constant
Measuring range
Operating temp.
Body materials
Electrode material
Temperature sensor
Min. immersion depth
Applications
Aqueous samples in general.
Limits
Conductivities of over 200 mS/cm.
Samples containing HF.
Warning.
The measuring range is only guaranteed when working
with CRISON conductivity meters
Warranty
CRISON guarantees this cell against manufacturing de-
fects.
Validity: 6 months from shipment date.
Limitations:
• Accidental damage.
• Inadequate applications.
• Non-fulfilment of the recommendations.
• Normal wear and tear of the cell.
QuALIty CERtIfICAtE
The attached cell 50 70 has passed the quality test and
fulfils the following specifications:
• EC error 5 % between 0.2 µS/cm...200 mS/cm.
• Temperature error 0.5 ºC.
0,7 cm
-1
0,2 µS/cm... 200 mS/cm
-30... 85 °C
glass
platinum
Pt 1000