5. Commissioning, operation

5.4 Performing measurements

1. Press "Measurement".
2. Configure the measuring methods and press "Start" to confirm.
The measurement begins.
Interim pumping of the gas back into the same gas compartment between
two measurements can cause a change in the gas quality. With several
measurements with interim pumping back, the measuring results with
respect to moisture of the gases can deviate from each other. For an optimal
measuring result, the gas should only be returned to the gas compartment
after all measurement processes have been completed.
WIKA operating instructions analytic instrument, model GA11
Measurement
Data
management
Settings