Performing A Control Test - Accu-Chek Performa 47949 Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Performing a Control Test

You need the meter, a test strip, and control solution Level 1 and/or Level 2.
The control level is printed on the bottle label.
1. Insert a test strip into the
meter in the direction of the
arrows. The meter turns on.
All manuals and user guides at all-guides.com
2. Make sure the code number
on the display matches the
code number on the test strip
container. If you miss seeing
the code number, take out the
test strip and reinsert it into
the meter.
OR
3. Select the control solution you
want to test. You will enter the
level later in the test.
49

Publicidad

Tabla de contenido
loading

Tabla de contenido