17
Technical Information
RS 232/RS 485 Communication Description
206
210
300
301
302
320
to
399
400
402
404
500
503
504
198/224
All manuals and user guides at all-guides.com
R
N
Error code
R
N
Actual rotation speed
(see note
in Hz
3)
R
N
Cycle time in minutes
(zeroed by the reset
command)
R
N
Cycle number (zeroed
by the reset command)
R
N
Pump life in hours
(zeroed by the reset
command)
Reserved to Agilent service
R
A
CRC EPROM (QE)
R
A
CRC Param. (PA)
R
A
CRC Parameter
structure
Reserved to Agilent service
R/W
N
RS 485 address
R/W
L
Serial type select
TV 301-P Navigator Controller User Manual / 87-901-001-01 (B)
Waiting intlk = 1
Starting = 2
Auto-tuning = 3
Braking = 4
Normal = 5
Fail = 6
Bit description: see the
following figure
100 to 1100
0 to 999999
0 to 9999
0 to 999999
QE8XXXX (where "XXXX"
are variable)
PA8XXXX (where "XXXX"
are variable)
"XXXX"
0 to 31
(default = 0)
0 = RS 232
1 = RS 485
(default = 0)