P P r r e e v v e e n n t t i i v v e e m m a a i i n n t t e e n n a a n n c c e e
Remove product from service before you perform the preventive maintenance inspection. Check all items listed during
annual preventive maintenance for all Stryker Medical products. You may need to perform preventive maintenance checks
more often based on your level of product usage. Service only by qualified personnel.
N N o o t t e e - - Clean and disinfect the exterior of the support surface before inspection, if applicable.
O O p p e e r r a a t t i i o o n n
Cleaning and disinfecting
Inspection
S S t t a a i i r r - - T T R R E E A A D D system belt
reconditioning (Model 6252)
Inspect the following items:
All fasteners secure (reference all assembly drawings)
All welds intact, not cracked or broken
No bent or broken tubing or sheet metal
No debris in wheels
All wheels secure and rolling
Front casters secure, rolling, and swiveling (if equipped)
Wheel locks hold wheels securely when on and clear the wheels when off
Chair unfolds and locks properly
No rips or cracks in seat or backrest
Restraint straps intact and working
Foot end carrying handles extend and lock in place
Head end carrying handles fold and unfold
Upper control handle extends and locks in all positions
S S t t a a i i r r - - T T R R E E A A D D system mechanism unfolds and locks in place (if equipped)
S S t t a a i i r r - - T T R R E E A A D D system belt rolls (if equipped)
S S t t a a i i r r - - T T R R E E A A D D system belt inner cords not showing; replace if necessary (if equipped)
S S t t a a i i r r - - T T R R E E A A D D system performs as desired; recondition belts if necessary (if equipped)
No lubricants present on the S S t t a a i i r r - - T T R R E E A A D D system belts or the track frame surfaces
Upper release handle cable not worn or frayed; replace if necessary (Model 6252)
Optional accessories and parts intact and operating properly
6252-009-005 Rev G.0
S S c c h h e e d d u u l l e e
Each use
For 1-25 calls per month, inspect chair
every 6 months
For 26-200 calls per month, inspect
chair every 3 months
For 201+ calls per month, inspect
chair monthly
After usage on approximately 500
flights of stairs or if the S S t t a a i i r r - - T T R R E E A A D D
system performance is sluggish
P P r r o o c c e e d d u u r r e e
See Cleaning in the Operations
Manual
See checklist below
See Track belt reconditioning in the
Maintenance Manual
21
EN