Descargar Imprimir esta página

Technical Data - wtw pH 330i Instrucciones De Operación

Ocultar thumbs Ver también para pH 330i:

Publicidad

Idiomas disponibles

Idiomas disponibles

Technical data

Dimensions
and weight
Mechanical
structure
Electrical safety
Test certificates
Ambient
conditions
pH/ORP
measuring ranges
Precision of
pH/ORP
(± 1 digit)
14
All manuals and user guides at all-guides.com
Technical data
Length [mm]
Width [mm]
Height [mm]
Weight [kg]
Type of protection
Protective class
cETLus, CE
Storage
Operation
Climatic class
pH
U [mV]
T [°C]
pH
(± 2 pH units from the
calibration point)
U [mV]
T [°C]
172
80
37
approx. 0.3
IP 66
III
- 25 °C ... + 65 °C
-10 °C ... + 55 °C
2
Measuring range
Resolution
- 2.000 ... + 19.999
0.001
- 2.00 ... + 19.99
0.01
- 999.9 ... + 999.9
0.1
- 1999 ... + 1999
1
- 5.0 ... + 105.0
0.1
± 0.003 at + 15 °C ... + 35 °C
± 0.01
± 0.2 at + 15 °C ... + 35 °C
NTC 30:
± 0.1
PT 1000:
± 0.5 at 0 °C ... 15 °C
± 0.1 at 15 °C ... 35 °C
± 1 at 35 °C ... 55 °C
pH 330i/340i

Publicidad

loading

Este manual también es adecuado para:

Ph 340i