ERKA. ERKAMETER 125 PRO Manual Del Usuario página 41

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

Compliance level
± 6 kV contact discharge
± 8 kV air discharge
± 2 kV for mains cables
± 1 kV for input and output
cables
± 1 kV common mode voltage
± 2 kV normal mode voltage
<5% UT for 0.5 periods
(> 95% drop)
40% UT for 5 periods
(> 60% drop)
70% UT for 25 periods
(> 30% drop)
< 5% UT for 5s (> 95% drop)
3 A/m
ERKA_Manual_ERKA125_PRO_11_2020.indd 41
ERKA_Manual_ERKA125_PRO_11_2020.indd 41
Electromagnetic environment
guidelines
Floors should consist of wood or
concrete or ceramic tiles must
be installed. If the floor consists
of synthetic material, the relative
humidity must amount to at least
30%.
The quality of the supply voltage
must correspond to that of a
typical commercial or hospital
environment
The quality of the supply voltage
must correspond to that of a
typical commercial or hospital
environment.
The quality of the supply voltage
must correspond to that of a
typical commercial or hospital
environment.
Magnetic fields for the mains
frequency should correspond to
the typical values found in a com-
mercial or hospital environment.
41
26.11.2020 10:56:54
26.11.2020 10:56:54

Publicidad

Tabla de contenido
loading

Tabla de contenido