Technical specifications for Electrical Muscle Stimulation
(EMS) mode
Waveform
Pulse amplitude Adjustable, 0~105 mA at 1,000 ohm Load each
Pulse Width
Pulse Rate
Contraction time Adjustable, 1—60 seconds, I Sec./step
Relaxation
(OFF) time
Ramp time
Synchronous (S) Stimulation of both channels occurs
Alternate (A)
Delay (D)
All manuals and user guides at all-guides.com
Onda de pulso cuadrada monofásica.
channel, 1 mA/step.
Adjustable, from 50 – 300µs.
Adjustable, from 1 – 150 Hz, 1 Hz/step.
Adjustable, 0—60 seconds, 1 Sec./step
Adjustable, 1 – 6 seconds, I Sec./step,
The "On" time will increase and decrease
in the setting value..
synchronously. The "ON" time including
"Contraction", "Ramp Up" and "Ramp Down" time.
ON TIME = Contraction + Ramp up + Ramp down
The Stimulation of the CH2 will occur after the
1st operation of CH1 is completed. In this
program, The "ON" time including "Contraction",
"Ramp Up" and "Ramp Down" time. The OFF
Time should be equal or more than the ON Time
ON TIME = Contraction + Ramp up + Ramp down
OFF TIME ≤ ON TIME
The Stimulation of the CH2 will occur after CH1
is started + Delay Time. In this program, The
"ON" time includes "Contraction", "Ramp Up" and
Ramp up + Ramp down Time. The OFF Time
should be equal or more than the ON Time +
Delay Time. Time Delay time is adjustable from
1 – 10 seconds in this program.